Aseptic Isolator

  • Isolator is used for sterility test. It is a highly airtight structure and has a high tightness. It corresponds to hydrogen peroxide sterilization. Air sampler etc can also be attached.

  • It is high airtight structure. The leak rate is less than 0.5% vol / h.
  • It corresponds to hydrogen peroxide sterilization.
  • It also corresponds to special dimensions. Also, various options such as air sampler can be included.
  • We also manufacture isolators for containment (negative pressure).

 

Cleanliness
main room: ISO Class5 (Class100), Subchamber: Progress (Progress by Supplying Clean N 2)
Dust collect efficiency
 More than 99.99%at 0.3μm (PAO)
Filter
HEPA filter
Processing air volume
Abt 30m³/min
Blowing wind speed
abt 0.35m/sec
Airflow condition
Down flow (only for the main room)
Residual oxygen concentration
Window part: made of polycarbonate, glove: made of butyl rubber (flat type), 0.1% or less (1000 ppm or less)
Internal pressure
Positive pressure specification (approx. + 40 Pa)
Nitrogen related
Supply nitrogen purity 99.999% or more, maximum 300 l / min
Nitrogen evacuation
Maximum 1 m³ / min × 2 systems Φ 50 duct exhaust
Power supply
AC200V 1Φ 60Hz
Equipment power
Abt 3000VA
Weight
Main room: about 450 kg, subchamber 150 kg
Structure
Body: SUS 304 # 400 polished (corner portion R )

 