-
Isolator is used for sterility test. It is a highly airtight structure and has a high tightness. It corresponds to hydrogen peroxide sterilization. Air sampler etc can also be attached.
Isolator is used for sterility test. It is a highly airtight structure and has a high tightness. It corresponds to hydrogen peroxide sterilization. Air sampler etc can also be attached.
Cleanliness |
main room: ISO Class5 (Class100), Subchamber: Progress (Progress by Supplying Clean N 2)
|
---|---|
Dust collect efficiency |
More than 99.99%at 0.3μm (PAO)
|
Filter |
HEPA filter
|
Processing air volume |
Abt 30m³/min
|
Blowing wind speed |
abt 0.35m/sec
|
Airflow condition |
Down flow (only for the main room)
|
Residual oxygen concentration |
Window part: made of polycarbonate, glove: made of butyl rubber (flat type), 0.1% or less (1000 ppm or less)
|
Internal pressure |
Positive pressure specification (approx. + 40 Pa)
|
Nitrogen related |
Supply nitrogen purity 99.999% or more, maximum 300 l / min
|
Nitrogen evacuation |
Maximum 1 m³ / min × 2 systems Φ 50 duct exhaust
|
Power supply |
AC200V 1Φ 60Hz
|
Equipment power |
Abt 3000VA
|
Weight |
Main room: about 450 kg, subchamber 150 kg
|
Structure |
Body: SUS 304 # 400 polished (corner portion R )
|
CÔNG TY CỔ PHẦN AIRTECH THẾ LONG
Trụ sở: Số 144, phố Việt Hưng, Tổ 3, P.Việt Hưng, Q. Long Biên, Tp.Hà Nội
Nhà máy 1: KCN Tân Quang, xã Tân Quang, H.Văn Lâm, Hưng Yên, Việt Nam
Nhà máy 2: Lô đất IN9.4, đường số 1, khu công nghiệp Cẩm Điền – Lương Điền, xã Cẩm Điền, huyện Cẩm Giàng, tỉnh Hải Dương, Việt Nam
Nhà máy 3: E5/34A, Đường Liên Ấp 5.6, xã Vĩnh Lộc A, huyện Bình Chánh, thành phố Hồ Chí Minh, Việt Nam