Phlegm collecting booth

It is a booth used at the time of sputum for tuberculosis examination, the interior is maintained at negative pressure. The exhaust from the booth is filtered with a HEPA filter and exhausted as sterile air.
Cough and sputum generated in the booth will not flow out to the outside due to negative pressure inside the booth.

 

  • In the booth, the negative pressure is maintained, there is no spill of cough etc. from the inside.
  • Air from the booth is filtered with a HEPA filter, sterilized and released.
  • When not in use, sterilize the inner surface with sterilizing light.
Model
ALF-902S
Dust collect efficiency
More than 99.99% at 0.3μm
Filter
HEPA filter
Air volume
Large airflow mode:abt 11.5m³/min、Silent mode:3.0m³/min *5 level switching
Power supply
AC100V 1φ 50/60Hz
Power consumption
Abt 150VA
Lighting
20W ×1
Sterilizing lamp
15W
Effective size
W890×D1085×H1920mm
Structure
Body, frame: Powder coated steel
Clean partition: Powder coated steel
Door: Aluminum frame, Floor: SUS 304
Option
Table SUS304

 

ALF-902S